NDIR310P Infrared gas analyzer

NDIR310 Pis a gas analyzer independently developed by our company for environmental monitoring and industrial field exhaust online analysis. The analyzer is mainly intended to measure concentrations of gases such as CO, CO2, N2O, CH4 and O2, and has the characteristics of high measurement accuracy, high stability and reliability, and short response time. The product is mainly used in the following fields:

Applications:

  • Continuous emission monitoring CEMS (analysis of CO, CO2, N2O, CH4, O2)
  • Continuous waste incineration flue gas emission monitoring (analysis of CO, CO2, O2)
  • Automotive exhaust emission monitoring (analysis of CO, CO2, HC, O2), etc.

Product Features:

  • Single beam dual wavelength technique, no filter wheel and other moving parts
  • Multiple-pass gas cell, low detection limit
  • High-precision temperature control, good environmental adaptability
  • MEMS IR source, electrical modulation
  • Modular design, high reliability, extensibility and maintainability

Technical Specifications:

Item

Value

Note

CO

(0~100)ppm

*Other ranges can be customized 

CO2

(0~25)%

*Other ranges can be customized 

N2O

(0~25)ppm

*Other ranges can be customized 

CH4

(0~500)ppm

*Other ranges can be customized 

HC

(0~500)ppm

*Other ranges can be customized 

O2

025%

 

Linearity

1% F.S.

 

Lack-of-fit

≤±2% F.S.

 

Repeatability

Less than 2%

 

Stability

Less than 2%

 

Sample Gas Flow rate

2~3)L/min

 

Zero Drift

±2%F.S.(24 hour)

@100ppm

Span Drift

±2%F.S.(24 hour)

 

Response Time

<60s

@2L/min

Influence of ambient temperature

Less than ±2%F.S.

 

Influence of voltage variations

Less than ±0.5%F.S.

 

Cross-sensitivity

CO2、CO <±1%F.S

 

CO2、N2O <±1%F.S

 

CO、N2O <±1%F.S

 

H2O、CO2 <0.2%

 

H2O、CO <0.5ppm

 

H2O、N2O <0.2ppm

 

H2O、CH4 <1ppm

 

Parallelism

≤2%

 

Operating Temperature Range

15~35°C

Analog I/O

6 channels with isolated 4-20mA output

 

Digital I/O

6 digital inputs (5-24V) 7 relay outputs

 

Serial port

1 RS232   1 RS485

 

Ambient Humidity

≤98%

 

Power Requirements

220VAC±10%,100W

 

Warm-up time

90-120min

 

Physical Dimensions

480mm(D)x306mm(W)133mm(H)

 

Weight

10KG