FTGAS-600
Product features
FTIR is short for Fourier Transform Infrared Spectroscopy,The basic principle is to transform the sequential interference signal generated by interferometer into spectral signal by Fourier transform.
The Michelson interferometer is a device that can divide a beam of radiation into two paths and then recombine the two beams after a path difference has been introduced. The variation of intensity of the beam emerging from the interferometer is measured as a function of path difference by a detector, which is called interferogram. ultimately yields the spectral by FFT algorithmic.
Components |
Range |
Extended Ranges |
CO |
(0~75)mg/m3 |
(0~2000)mg/m3 or higher |
CO2 |
(0~25)% |
(0~50)% |
CH4 |
(0~50)mg/m3 |
(0~5000)mg/m3 or higher |
NO |
(0~150)mg/m3 |
(0~2000)mg/m3 or higher |
NO2 |
(0~150)mg/m3 |
(0~2000)mg/m3 or higher |
N2O |
(0~50)mg/m3 |
(0~2000)mg/m3 or higher |
NH3 |
(0~50)mg/m3 |
(0~200)mg/m3 or higher |
SO2 |
(0~50)mg/m3 |
(0~2000)mg/m3 or higher |
HCl |
(0~75)mg/m3 |
(0~200)mg/m3 or higher |
H2O |
(0~40)% |
|
HF |
(0~50)mg/m3 |
(0~200)mg/m3 or higher |
System |
|
Physical Dimensions |
435mm(D) x175mm(H) x435(D) |
Weight |
23Kg |
Power Requirements |
220VAC±10%/Max 2500W Ave 150W@ AT 25℃ |
Gas Temperature |
Up to 180℃ |
Operating Temperature Range |
15-35℃ |
Purge and Zero gas |
Nitrogen(99.999% or higher recommended) |
Warm-up Time |
2 hours |
Sample Gas Pressure |
Ambient |
Sample Gas Flow rate |
2 – 3 l/min recommended |
Sample Gas Filtration |
Minimum 2 µm particulate filtration |
Gas fittings |
Sample in: 6 mm Swagelok, stainless steel |
Sample out:8 mm Swagelok, stainless steel |
|
Interferometer purge: 6 mm Swagelok stainless steel |
|
Output Port |
Ethernet port |
Communication Protocols |
MODBUS,OPCUA |
Measuring principle |
FTIR |
Performance |
|
Zero Drift |
< 2 %@24 hours |
Span Drift |
< 1 %@24 hours |
Response Time |
Typically <100s@3L/min |
Linearity |
±1%F.S. |
Repeatability |
1%. |
Detection Limit |
<1% F.S. |